Menu
It is important to obtain representative histological samples of solid biliopancreatic lesions without a clear indication for resection. The role of new needles in such task is yet to be determined.
To compare performance assessment between 20G double fine needle biopsy (FNB) and conventional 22G fine needle aspiration (FNA) needles for endoscopic ultrasound (EUS)-guided biopsy.
This prospective study examined 20 patients who underwent the random puncture of solid pancreatic lesions with both needles and the analysis of tissue samples by a single pathologist.
The ProCore 20G FNB needle provided more adequate tissue samples (16 vs. 9, p=0.039) with better cellularity quantitative scores (11 vs. 5, p=0.002) and larger diameter of the histological sample (1.51±1.3 mm vs. 0.94±0.55 mm, p=0.032) than the 22G needle. The technical success, puncture difficulty, and sample bleeding were similar between groups. The sensitivity, specificity, and diagnostic accuracy were 88.9%, 100%, and 90% and 77.8%, 100%, and 78.9% for the 20G and 22G needles, respectively.
The samples obtained with the ProCore 20G FNB showed better histological parameters; although there was no difference in the diagnostic performance between the two needles, these findings may improve pathologist performance.
Desenvolvido por Surya MKT